Refraction Unit Delivery and Installation: What to Prepare
Our refraction units usually ship by sea freight because of the size and weight of the furniture. They arrive close to ready to use, and the customer installs them with remote support from ENOCH. This guide lists what to prepare before the unit arrives, using the figures each model’s page publishes.
Where a figure is not published for a model, this guide says so; ask us for it before you rely on it.
The instruments you mount on the unit are not part of the furniture supply unless a quotation states otherwise.
Five things to prepare.
1. Power
Each unit is specified for AC 220 V / 50 Hz or 110 V / 60 Hz; the version is specified on the quotation. Listed power runs from 3 W (C-3, C-10A, C-800) and 20 W (HRT-1A, HRT-3) to 50 W (MI-180) and 100 W (TG-510, TG-1100A). The TR-500A lists a 55 W motor and a 40 W illuminator, the TR-900 a 55 W motor and a 16 W illuminator. The instruments you mount need their own supply; the TG-1010 and TG-510 add 3 V / 6 V / 12 V AC outlets.
2. Floor area
Some pages give a unit size and a larger installation footprint. The HRT-3 is 1650 × 1050 mm but needs a 1500 × 2500 mm footprint; the C-10A is 1050 × 1300 mm with a 1300 × 1300 mm footprint. Plan to the footprint where one is listed. Leave space where the chair rotates or reclines: the TR-500A page notes that a reclining backrest needs room behind the chair, while the fixed-backrest version can stand close to a wall.
3. Ceiling height
Listed overall heights run from 1550 mm (C-10A, C-12) to 2130 mm (TR-500A). The TG-1010 is 2090 mm. The TR-500A and TR-900 publish the same working geometry and differ in envelope: 2130 mm against 1990 mm. The TR-900 page recommends it where overall height is the constraint.
4. Moving it in
Weights and packing sizes are published for some models only: the TG-510 lists packing of 1120 × 1120 × 1100 mm and 200 kg gross, the TG-1100A 137 kg gross, the TR-500A 168 kg and the TR-900 175 kg machine weight. For the other models, ask for the packing list before you measure doors, lifts and stairs.
5. After arrival
The customer installs, with remote support from ENOCH. The standard warranty is 12 months from delivery; within that period ENOCH supplies the replacement parts needed for any fault not caused by misuse or accidental damage.
Space, power and weight across our units.
| Model | Space (mm) | Power as listed | Weight or packing as listed |
|---|---|---|---|
| C-3 | 940 × 800 × 1850 | 3 W | not listed |
| C-10A | Unit 1050 × 1300 × 1550; footprint 1300 × 1300 | 3 W | not listed |
| C-800 | 2100 × 1200 × 1700 | 3 W | not listed |
| HRT-1A | (1470–1620) × (1020–1360) × 1900 | 20 W | not listed |
| HRT-3 | Unit 1650 × 1050 × 1900; footprint 1500 × 2500 | 20 W | not listed |
| MI-180 | 900 × 1000 × 1700, space dimensions | 50 W total | not listed |
| TG-510 | 1750 × 1600 × 2060 | 100 W total | Packing 1120 × 1120 × 1100 mm; 200 kg gross |
| TG-1100A | 1570 × 1930 × 1730 | 100 W total | 137 kg gross |
| TR-500A | 1510 × 1000 × 2130 | Motor 55 W; illuminator 40 W | 168 kg machine weight |
| TR-900 | 1500 × 1000 × 1990 | Motor 55 W; illuminator 16 W | 175 kg machine weight |
All ten are specified for AC 220 V / 50 Hz or 110 V / 60 Hz. The C-3, C-10A and C-800 list their figure as power / lighting. Figures are reproduced as each model’s specification gives them; where a figure is not listed, ask us before you rely on it.
What to tell us before ordering.
Send us the destination, the quantity, the room dimensions and ceiling height, and the route the unit must take to the room: doors, lifts and stairs. Tell us your supply voltage (220 V / 50 Hz or 110 V / 60 Hz) and which instruments you will mount.
We reply with a written configuration, lead time and quotation, usually within one business day. The plug type and voltage version are confirmed against the quotation.
Questions about delivery and installation.
How are refraction units delivered?
Usually by sea freight, because of the size and weight of the furniture. The units arrive close to ready to use.
Who installs the unit?
The customer, with remote support from ENOCH.
Is there a warranty?
The standard warranty is 12 months from delivery. Within that period ENOCH supplies the replacement parts needed for any fault not caused by misuse or accidental damage.
What voltage do the units need?
Each is specified for AC 220 V / 50 Hz or 110 V / 60 Hz. The version you need is specified on the quotation.
How heavy is a refraction unit?
Weight is published for only some models: 200 kg gross for the TG-510, 137 kg gross for the TG-1100A, 168 kg for the TR-500A and 175 kg for the TR-900. Ask us for the others.
Are the instruments included?
No. Phoropters, chart projectors and autorefractors shown in photographs illustrate use and are not part of the furniture supply unless a quotation states otherwise.
Related guides and equipment.
Tell us the room and the route in.
Send your dimensions, quantity and destination. We reply with a configuration, lead time and quotation.